back
EP-INFR1200 - Infrared Temperature Calibrator
Article-No.: EP-INFR1200manufacturer
Temperature Range: 300 °C to 1200 °C
Measuring Zone Diameter: 38 mm
Emissivity: 0.99 ± 0.005
Uncertainty: 2 K + 0.0025 · T [°C]
Reproductivity: 0.5 K + 0.001 · T [°C]
Stability: 0.3 K
Resolution: 0.1 K
Uniformity: 0.3 K + 0.001 · T [°C] (for central area ∅ 20 mm)
Heating Time: to 300 °C = 30 min, to 1200 °C = 50 min
Temperature Sensor: Type R PT13Rh
Measuring Zone Diameter: 38 mm
Emissivity: 0.99 ± 0.005
Uncertainty: 2 K + 0.0025 · T [°C]
Reproductivity: 0.5 K + 0.001 · T [°C]
Stability: 0.3 K
Resolution: 0.1 K
Uniformity: 0.3 K + 0.001 · T [°C] (for central area ∅ 20 mm)
Heating Time: to 300 °C = 30 min, to 1200 °C = 50 min
Temperature Sensor: Type R PT13Rh
The EP-INFR1200 calibration source is used to calibrate and check radiation thermometers and thermal imaging cameras in the temperature range from 300 °C to 1200 °C. The design as a cavity radiator guarantees a very high emissivity.
The temperature setting can be done directly at the radiator or via the serial interface RS-485 (optional USB adapter) using a PC. A digital PID controller and a precision temperature sensor guarantee outstanding accuracy and stability.
The temperature setting can be done directly at the radiator or via the serial interface RS-485 (optional USB adapter) using a PC. A digital PID controller and a precision temperature sensor guarantee outstanding accuracy and stability.